Reports until 18:08, Friday 01 March 2024
H1 CAL
anthony.sanchez@LIGO.ORG - posted 18:08, Friday 01 March 2024 (76079)
PCAL EX End Station Measurement & Lab Measurements

Post EX Vent PCAL EndX Station Measurement
A rare Non Tuesday PCAL End station measurement was done on leap day Feb 29th 2024, the PCAL team(    
Francisco L., Julian G, & Tony S.) went to EndX with Working Standard Hanford aka WSH(PS4) and took an End station measurements.
The EndX Station Measurement was carried out according to the procedure outlined in Document LIGO-T1500062-v15, Pcal End Station Power Sensor Responsivity Ratio Measurements: Procedures and Log, and was completed by 11:45 am.
Note:
After the normal measurement.

Measurement Log

First thing we did is take a picture of the beam spot before anything is touched!
We expected the beams to move due to the the vent, and it does seem like both beams have moved. It may be neccisary for us to adjuct the beams. I believe the distance between the holes on this target is 5mm center on center.

Martel:
Martel Voltage source applies voltage into the PCAL Chassis's Input 1 channel. We record the GPStimes that a -4.000V, -2.000V and a 0.000V voltage was applied to the Channel. This can be seen in Martel_Voltage_Test.png. We also did a measurement of the Martel's voltages in the PCAL lab to calculate the ADC conversion factor, which is included on the above document.

Plots while the Working Standard(PS4) is in the Transmitter Module during Inner beam being blocked, then the outer beam being block, followed by the background measurment: WS_at_TX.png.

The Inner, outer, and background measurement while WS in the Receiver Module: WS_at_RX.png.

The Inner, outer, and background measurement while RX Sphere is in the RX enclosure, which is our nominal set up without the WS in the beam path at all.:  TX_RX.png.

The last picture is of the Beam spots after we had finished the measurement. I did not attempt to adjust the beam positions.


All of this data is then used to generate LHO_EndX_PD_ReportV2.pdf which is attached, and a work in progress in the form of a living document.


All of this data and Analysis has been commited to the SVN :
https://svn.ligo.caltech.edu/svn/aligocalibration/trunk/Projects/PhotonCalibrator/measurements/LHO_EndX/

 

PCAL Lab Responsivity Ratio Measurement:
A WSH/GSHL (PS4/PS5)FrontBack Responsivity Ratio Measurement was ran, analyzed, and pushed to the SVN.
The analysis of this measurement produces 4 PDF files which we use to vet the data for problems.
raw_voltages.pdf
avg_voltages.pdf
raw_ratios.pdf
avg_ratios.pdf

 


Obligitory BackFront PS4/PS5 Responsivity Ratio:
PCAL Lab Responsivity Ratio Measurement:
A WSH/GSHL (PS4/PS5)BF Responsivity Ratio measurement was ran, analyzed, and pushed to the SVN.
The analysis of this measurement produces 4 PDF files which we use to vet the data for problems.
raw_voltages2.pdf
avg_voltages2.pdf
raw_ratios2.pdf
avg_ratios2.pdf

Alphatrends

This adventure has been brought to you by Francisco, Julian, & Tony Sanchez.

 

 