OMC alignments comparison between yesterday and this morning

Trent Matt and Jennie

We looked at the pitch and yaw of the IFO output mirrors (OM1-3+C) this morning around 6am and the same pitch and yaw channels at 6am yesterday; the only one that we saw with significant differences were OM2 Pitch and Yaw, which both went down about 30 and 20 microradians (we think this is the unit counts represents), respectively.