Reports until 14:58, Saturday 09 March 2024
H1 ISC
stefan.ballmer@LIGO.ORG - posted 14:58, Saturday 09 March 2024 - last comment - 14:42, Monday 11 March 2024(76228)
DCPA Anti-TIA (Anti-Transimpedance Amplifier Filters) show small missmatch

Meashured the transfer function between DCPD_A and DCPD_B usinng a broadband injection. There is a slight frequency-dependent disagreement (0.2dB) between the two diodes, right around 25Hz, where the TIA has its complex poles.

It looks like  one or both Anti-TIA filters are slightly mismateched. We won't tweak it today because we don't know with one is the culprit.

 

 

Images attached to this report
Comments related to this report
louis.dartez@LIGO.ORG - 11:46, Monday 11 March 2024 (76258)
this looks to be related to the TIA response drift reported in LHO:75986.
stefan.ballmer@LIGO.ORG - 14:42, Monday 11 March 2024 (76266)

The template used for this measurement can be found here:

/ligo/home/controls/sballmer/20240311/AoverBmeasurement.xml

Note: the max peak-peak deviation between DCPD_A and DCPD_B is 2.5%.

Images attached to this comment