Displaying report 1-1 of 1.
Reports until 15:51, Tuesday 05 March 2024
H1 ISC
minhyo.kim@LIGO.ORG - posted 15:51, Tuesday 05 March 2024 - last comment - 14:21, Monday 11 March 2024(76131)
OMC ASC signal comparison
From Minhyo & Preet 
After changing of OMC control, it might be good to check the signals.
(Refer to Jennie's alog: https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=76126, https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=76103)

Figures are comparison between Mar. 5 and one of the date in O4a (Jan. 2. 2024): 
1) SUS-OM3_LOCK comparison with both PIT and YAW
2) SUS-OMC_ASC comparison with both PIT and YAW
3) OMC-ASC_QPD A comparison with both PIT and YAW
4) OMC-ASC_QPD B comparison with both PIT and YAW

Seems that the signal went up in high frequency (4 ~ 10 Hz), which is pretty similar with QPD signals.
Images attached to this report
Comments related to this report
minhyo.kim@LIGO.ORG - 14:21, Monday 11 March 2024 (76264)
Update on Mar. 11

After locking of detector, compared the signals between O4a (Jan. 2, 2024, 05:00:00 UTC) and last Sat. (Mar. 9, 2024, 10:00:00 UTC).

1) SUS-OM3_LOCK comparison with both PIT and YAW
2) SUS-OMC_ASC comparison with both PIT and YAW
3) OMC-ASC_QPD A comparison with both PIT and YAW
4) OMC-ASC_QPD B comparison with both PIT and YAW

QPD signals went down from the last time (Mar. 5), which is similar to SUS (OM3, OMC) signals.
Images attached to this comment
Displaying report 1-1 of 1.