E. Hall, S. Pandey, S. Dwyer, L. Dartez
We took another look at the new DARM configuration this evening. Evan had a suggestion for a new test: move into NLN_ETMY, pause here before switching back to ETMX in the new configuration (but after all the new filters have been put into place) to take measurements of the new ETMX OLG before activating it. I took us into NLN_ETMY using the guardian then manually ran the instructions in NEW_DARM to set up the ETMX locking filters but paused before switching arms (so we were still locked with ETMY, notably without the MICH feedforward which remained inactive for the rest of our testing).
We measured H1:SUS-ETMX_L3_LOCK_L_IN1 / H1:SUS-ETMX_L3_LOCK_L_IN2 and H1:SUS-ETMY_L3_LOCK_L_IN1 / H1:SUS-ETMY_L3_LOCK_L_IN2 (DTT template paths are below) to isolate the OLG for the new DARM path on the x-arm. Some post-processing of the data is needed to do this. We'll follow up with that.
Finishing the Transition to the new state:
After the measurements above (but before we had a chance to look at the data) we tried slowly bringing up ETMX slowly starting with an L3 LOCK gain of 0.01 (and an ETMY L3 LOCK gain of 1). We immediately lost lock. This indicates that there is something severely wrong with the new DARM configuration as it's currently installed. Most likely, this is not something as subtle as an instability. We'll need to walk through these steps more carefully next time.
DTT Templates:
H1:SUS-ETMX_L3_LOCK_L_IN1 / H1:SUS-ETMX_L3_LOCK_L_IN2: /ligo/home/louis.dartez/projects/20240311_new_darm_investigations/SUSETMX_L3_SS_new_darm.xml
H1:SUS-ETMY_L3_LOCK_L_IN1 / H1:SUS-ETMY_L3_LOCK_L_IN2: /ligo/home/louis.dartez/projects/20240311_new_darm_investigations/SUSETMY_L3_SS_new_darm.xml
Relevant logs:
- Old Matlab model revived (LHO:75584)
- Unsuccessful attempts in early 2024 (LHO:75308)
- Times spent successfully in new DARM state (LHO:75631)
- Calibration attempts in new DARM state (LHO:74977)