Displaying report 1-1 of 1.
Reports until 15:57, Tuesday 12 March 2024
H1 CDS
david.barker@LIGO.ORG - posted 15:57, Tuesday 12 March 2024 (76304)
CDS Maintenance Summary: Tuesday 12th March 2024

WP11766 Shorten SWWD SUS-EY trip time 
Dave:
The h1iopsusey model was changed to reduce the SWWD SUS timer from 900s to 600s. This means that the SUS DACKILL time to trip is shortened from 20mins to 15mins and is no longer coincident with the HWWD tripping. 

This IOP restart required the restart of all the models on h1susey, and was done in conjuction with the restart of h1susetmy.

After all the models had been running again for a few minutes, h1susey locked up. It was not on the network (no ping, ssh). We fenced it from Dolphin and power cycled it via its IPMI management port.

WP11743 SUS DACKILL Removal
Jeff, Oli, Jim, Dave:
Today we removed the SUS DACKILL from the end station models h1susetmx and h1susetmy. The dolphin ipc receiving models h1isietmx, h1isietmy were modified to replicate a good ipc receiver.
In addition the h1susetmxpi and h1susetmypi models also needed replicated receivers.


Unfortunately I got the ERR replication value incorrect, it should be 0 and not 1. The receiver models were restarted with the correct values.

The first round of restarts required a DAQ restart, the subsequent changes only needed model restarts.

The TMS models were modified later, requiring a second DAQ restart.

WP11761 sw-fces-cds0 Firmware Upgrade
Jonathan:
The firmware on the FCES switch sw-fces-cds0 was upgraded.

WP11765 DMT Upgrade and Reboots
Dan:
This action was canceled for today and will be scheduled for a later maintenance period.

DAQ Restart
Jonathan, Erik, Dave:

The DAQ was restarted partially once, completely twice for the above model changes.

The first restart (partial 0-leg only) at 12:30 highlighted that the PI models were showing IPC Rx errors due to a removed SUS SHMEM channel.

The second full restart at 13:06/13:12 was in support of the SUS-ETM, SUS-ETM-PI and ISI-ETM model changes.

the third full restart at 14:47/14:52 was in support of the SUS-TMS model changes.

There were no major issues with the DAQ restarts, just the usual GDS second restarts for channel
list synchronization.

Restart/Reboot Log

Tue12Mar2024
LOC TIME HOSTNAME     MODEL/REBOOT
12:14:27 h1seiex      h1isietmx <<< SUS-ETM and ISI-ETM model restarts
12:14:58 h1susex      h1susetmx
12:15:32 h1seiey      h1isietmy
12:16:47 h1susey      h1iopsusey <<< EY IOP and all Models restarts
12:17:04 h1susey      h1susetmy
12:17:18 h1susey      h1sustmsy
12:17:32 h1susey      h1susetmypi


12:24:21 h1susey      ***REBOOT*** <<< h1susey locked up, needed a reboot
12:26:25 h1susey      h1iopsusey
12:26:38 h1susey      h1susetmy
12:26:51 h1susey      h1sustmsy
12:27:04 h1susey      h1susetmypi


12:30:40 h1daqdc0     [DAQ] <<< Partial 0-leg restart
12:30:49 h1daqfw0     [DAQ]
12:30:49 h1daqtw0     [DAQ]
12:30:50 h1daqnds0    [DAQ]
12:30:58 h1daqgds0    [DAQ]
12:34:33 h1daqgds0    [DAQ]


12:50:53 h1susex      h1susetmxpi <<< PI model change, remove IPC Rx
12:55:45 h1susey      h1susetmypi


13:06:50 h1daqdc0     [DAQ] <<< Full DAQ restart for susetm, susetmpi, isietm models
13:07:01 h1daqfw0     [DAQ]
13:07:01 h1daqtw0     [DAQ]
13:07:02 h1daqnds0    [DAQ]
13:07:10 h1daqgds0    [DAQ]
13:08:42 h1susey      h1susetmypi <<< Problem is etmypi build (install too quick), rebuild and install.
13:12:08 h1daqdc1     [DAQ]
13:12:20 h1daqfw1     [DAQ]
13:12:20 h1daqtw1     [DAQ]
13:12:22 h1daqnds1    [DAQ]
13:12:30 h1daqgds1    [DAQ]
13:12:56 h1daqgds1    [DAQ]


13:49:38 h1seiex      h1isietmx << Fix the replicated IPC ERR=0
13:55:49 h1seiey      h1isietmy
13:58:35 h1susex      h1susetmxpi
14:01:46 h1susey      h1susetmypi


14:43:40 h1susex      h1sustmsx <<< New TMS models
14:44:06 h1susey      h1sustmsy


14:47:04 h1daqdc0     [DAQ]  <<< DAQ restart for TMS model changes
14:47:15 h1daqfw0     [DAQ]
14:47:16 h1daqnds0    [DAQ]
14:47:16 h1daqtw0     [DAQ]
14:47:24 h1daqgds0    [DAQ]
14:47:50 h1daqgds0    [DAQ]
14:52:21 h1daqdc1     [DAQ]
14:52:33 h1daqfw1     [DAQ]
14:52:33 h1daqtw1     [DAQ]
14:52:34 h1daqnds1    [DAQ]
14:52:42 h1daqgds1    [DAQ]

Displaying report 1-1 of 1.