louis.dartez@LIGO.ORG - posted 21:59, Thursday 14 March 2024 (76410)
NEW DARM guardian state bug fix
[Gabriele, Louis]

we added a counter increment at the end of the run method in the NEW_DARM state (ISC_LOCK). The missing self.counter += 1 was causing the state to continuously set LSC-MICHFF_GAIN and LSC-SRCLFF1_GAIN to 1. This incidentally caused at least one lockloss as Gabriele was trying to run a MICH feedforward measurement tonight.


this was on line 6207