The readback of the broadband PD in transmission of the SHG (SQZ-SHG_TRANS_DC) was electronicslly saturating. We switched the gain from 10x to 1x on the generic PD interface box and adjusted the transimpedance gain accordingly. The readback now shows ~1.2V and 7.2mW.