Reports until 11:49, Thursday 21 March 2024
H1 SQZ
camilla.compton@LIGO.ORG - posted 11:49, Thursday 21 March 2024 (76594)
PMC lock checker added to SQZ_MANAGER

Naoki, Camilla

When we got to NLN at 2024/03/21 17:38UTC the SQZ_PMC wasn't locked, SQZ manager didn't realize this and was still trying to lock the SHG and OPO. We've now added a @PMC_checker to SQZ_MANGER to check that the SQZ_PMC guardan is in LOCKED.