arnaud.pele@LIGO.ORG - posted 12:09, Friday 06 September 2013 (7662)
ITMX clear of rubbing
A spectra of ITMX has been taken this morning, and the frequencies of the peaks were compared with the ones from the "wire-hang model". They were matching very well, meaning the main chain of ITMX is currently freely suspended.