CLF second AOM saturation

As shown in the attachment, the output of CLF second AOM is 61dBm while the nominal is 31dBm. The output increased 10 days ago when new CLF VCO is installed in 76297.

Images attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 18:44, Friday 22 March 2024 (76646)

This readback is part of the CLF VCXO chassis. The new VCO doesn't have this feature, so the value is bogus.