TITLE: 03/24 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Earthquake
CURRENT ENVIRONMENT:
SEI_ENV state: SEISMON_ALERT
Wind: 5mph Gusts, 4mph 5min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.20 μm/s
QUICK SUMMARY:
H1:PEM-CS_DUST_PSL101 Error: data set contains 'not-a-number' (NaN) entries
Waited For ground motion to go down.
started Initial alignment and X arm looked a little strange so I took ISC_lock to GREEN_ARMS_MANUAL [13]
Which allowed me to touch up X arm a bit.
I ten too ISC_LOCK back to Initial_Alignment and this screen shot was taken where X arm seems to be "a fat noisey signal" more so than usual. I though that was a little weird.
COMM = +3
Diff= -8
Relocking got to Transition from ETMX then a lockloss 00:51 UTC .
Lockloss from Find IR
Lockloss from LOW_NOISE_LENGTH_CONTROL
Redoing Initial_Alignment
Another Lockloss at LOW_NOISE_LENGTH_CONTROL
None of the Locklosses at LOW_NOISE_LENGTH_CONTROL seem like they were caused by ground motion.
Incoming Earthquake from Timor_Leste 5.7
Lockloss at Find IR
Another Lockloss at Find IR
Did I just hear thunder? Do we have a lightning sensor to check out?
Checked the H0:FMC-EX_MAINS_CHAN_ voltage channels for spikes. I didn't see anything Like Oli's Alog.
I'm still not sure why we are losing lock at LOW_NOISE LENGTH CONTROL yet because we keep getting Locklosses at DRMI and below.
Holding for a few 5.7 Mag earthquakes to roll through.
ALS X had a PLL PDH error and gave me an ARM In fault code
Crystal Frequecy out of Tolerance
Restored all (except SQZ back to the point of Right After an the First Initial Alignment tonight that lead us to the Low Noise Length Control State. GPS Time: 1395274481
Then Touched up ALS X and Y tried to lock and couldnt get past PRMI. Going back to initial alignment again.