ryan.short@LIGO.ORG - posted 14:27, Monday 25 March 2024 (76690)
OMC and CALEY No Longer Excluded from DIAG_EXC

R. Short, T.J. Shaffer

We discovered this afternoon that starting an injection on the OMC model did NOT cause H1 to drop observing. After looking into the DIAG_EXC Guardian node, which checks for errant excitations and prevents going into observing accordingly, we discovered two models had been added to the exclude_list: omc and caley. These appear to have been added in May of 2023 (pre-O4a) and never committed to svn. TJ and I both recalled these models having ongoing excitations in the leadup to O4a, but neither of us could remember why. Clearly they are not needed anymore, so they have been removed, and the exclude_list is again empty (the one exception remains for the CAL-INJ_CW_EXC excitation).

Errant injections into the omc and caley models will now properly prevent H1 from observing.