Reports until 15:29, Monday 25 March 2024
H1 SQZ (SQZ)
julian.gurs@LIGO.ORG - posted 15:29, Monday 25 March 2024 (76691)
Spare SHG single pass temperature scan

A sinc function / temperature measurement from the single pass (spare) SHG is shown. The basic idea of the experiment was to see if the crystal is intact (it is). For this purpose, the crystal was pumped with 1064 nm and after one pass (single pass) it was checked how much green light was generated. The temperature was adjusted to find the best possible phase matching temperature.

No similarity to Mount Saint Helens, as described in 76239, was observed. Conclusion, the crystal is intact. (What seems strange is the asymmetry in the side maxima.)

Setup:

Images attached to this report