Evan, Naoki, Nutsinee, Camilla
Today Sheila added nodes.set_managed() in INIT state of SQZ_FC guardian and let SQZ_FC guardian go through the INIT state. However, the FC2 M1 offload filter in the INIT state was old filter setting in 68914, which caused GR_SUS_LOCKING failure. We reverted it as shown in the attachment and updated the FC2 M1 offload filter in INIT state.
Attached a screen shot of the old (wrong) settings on top, correct one at the bottom.
In 76503, the limiter for FC2 M1 was not in INIT state of SQZ_FC guardian, but we added the limiter.