Naoki, Vicky, Sheila, Camilla
To scan the sqz angle to optimize the squeezing in the bucket, we copied the SCAN_SQZANG state in SQZ_MANAGER guardian in LLO. This state will find the optimal sqz angle to minimize the BLRMS3 at 350 Hz. We replaced the 0.1 Hz LP with 1 Hz LP for BLRMS3. We will test this state tomorrow.
We tested the SCAN_SQZANG state and it seems to work. The result is saved here.
https://lhocds.ligo-wa.caltech.edu/exports/SQZ/GRD/SQZANG_SCAN/