Reports until 14:57, Monday 01 April 2024
H1 AOS
robert.schofield@LIGO.ORG - posted 14:57, Monday 01 April 2024 - last comment - 15:00, Monday 01 April 2024(76847)
Scattering noise from input arm vibration injections minimized with new setting of IRMY yaw offset

While studying the improvement made by stray light modifications over the break (summary alog soon), I found that the coupling was minimized by changing ITMY R0 yaw offset from -300 to -250. We want to change the settings before my alog, hence this mini-alog.

Comments related to this report
jenne.driggers@LIGO.ORG - 15:00, Monday 01 April 2024 (76848)

I have set and saved that value of -250 for the IY reaction mass yaw offset in both the safe and observe SDF files.  I did this right around when DRMI had just caught lock during this acquisition, so it seems fine (as we'd expect) for locking with this very slightly different offset.  Attached are the safe and observe snap screenshots of my accepting the new value.

Images attached to this comment