mark.barton@LIGO.ORG - posted 15:43, Monday 09 September 2013 - last comment - 16:38, Monday 09 September 2013(7688)
TMSX OSEM OLs, gains and offsets

Cheryl and Keita backed out the TMSX OSEMs and I logged the OL values and calculated gains and offsets using the Matlab script /ligo/svncommon/SusSvn/sus/trunk/Common/MatlabTools/prettyOSEMgains.m:

>> prettyOSEMgains('H1','TMSX')
M1F1 25441 1.179 -12720
M1F2 22328 1.344 -11164
M1F3 28300 1.060 -14150
M1LF 23159 1.295 -11580
M1RT 26172 1.146 -13086
M1SD 28743 1.044 -14372

I entered the new gains and offset and updated and commited the safe.snap.

Comments related to this report
keita.kawabe@LIGO.ORG - 16:38, Monday 09 September 2013 (7691)

And after the above was done OSEMs were centered such that the H1:SUS-TMSX_OSEMINF_??_OUT numbers are within +-20um.

SUS team was notified that TMSX is ready for SUS testing.