H1 SQZ (ISC)
victoriaa.xu@LIGO.ORG - posted 13:34, Tuesday 02 April 2024 - last comment - 10:25, Friday 05 April 2024(76885)
Unsqueezed DARM excess noise getting lower, more similar to O4a

Attachment 1 - Unsqueezed (left) vs. squeezed (right) DARMs in ER16, with respect to O4a. Like Gabriele's plots, the ratios on the bottom show a given trace / O4a reference, so ratio > 1 means more noise now than O4a.

Colors: Rainbow colors are ordered past to present: e.g. Red = O4a, ... Purple = now.

Attachment 2 - SQZ dBs after subtracting non-quantum noise, with various recent PSAMS settings. Upper subplot: Anti-SQZ, Lower subplot: SQZ. Lines are a moving average of the data. Some caveats for the subtraction: here I am using the same H1 unsqueezed quantum noise model for all subtractions; this may not be the most accurate, but I think it should be fairly close.

 

Reference alogs wth times that I used to grab GDS_STRAIN_CLEAN:

Images attached to this report
Comments related to this report
victoriaa.xu@LIGO.ORG - 10:25, Friday 05 April 2024 (76978)

Adding FDS times from the DARM comparison in 76955, and only including times with sqz/no-sqz comparable to those recent darm comparisons.

Attachment 1 - Unsqueezed (left) vs. squeezed (right) DARMs in ER16, with respect to O4a. Like Gabriele's plots, the ratios are trace / O4a reference, so ratio > 1 means more noise now than O4a.

Attachment 2 - SQZ dBs after subtracting non-quantum noise. Analysis with non-quantum noise subtraction for this week's PSAMS data in 76925 and 76949 ongoing.

Attachment 3 - Comparison to a very basic gwinc noise model (pulled from H1 NB) for squeezed darm. For example the dashed purple "Unsqueezed QN model" is (almost) what I am using for subtracting non-quantum noises.

Colors are rainbow-ordered time-wise.  Red = O4a reference.  Purple = 4/4 times from 76955 for FDS DARM comparisons.

 

Note: here the PSAMS settings are referenced in terms of strain voltages. Just pointing out that LLO has a guardian to servo PSAMS voltage to meet the "_TARGET" strain gauge voltage LLO:63403.

 

Reference alogs wth times used to grab GDS_STRAIN_CLEAN:

  • O4a references - sqz data set 74935, and Gabriele's times from e.g. 76825
  • 76434 - 3/17, PSAMS 120/120
  • 76540 - 3/20-3/25, PSAMS 150/90
          --> corresponding SQZ data set saw ~10-15% higher frequency-dependent losses at 100 Hz vs. kHz unlike earlier Mar16, possibly due to PSAMS settings following 76507
  • 7675776777 - 3/27-29, PSAMS 100/100, since SQZ ZM alignment guardian
  • 76872 - 4/2, PSAMS 100/100 - opportunistic NO SQZ ~1hour
  • 76955 - 4/4, DARM comparison
Images attached to this comment