https://ldas-jobs.ligo.caltech.edu/~gabriele.vajente/bruco_1396341217_GDS_CALIB_CLEAN/ DHARD_Y is pretty bad, much worse than usual. Probably need a retuning of A2L
Jennie, Sheila, Camilla
We expect DHARD yaw to be worse. In the seconds half of last night's lock the CAM_YAW1 offset was changed by ~7 counts, see 76961. It is expected that A2L would need to be retuned, we are still working on adding these 76961 changes to the guardian.
Jennie, Sheila, Camilla