04/07/2024 11:32:46UTC A dolphin crash (77040) caused the detector to unlock, tripped HAM1 Hepi, and caused a bunch of front ends to trip. Called Dave and Erik, who were able to fix everything
12:22 ISI ITMX Stage 2 watchdog tripped
I took IMC_LOCK to DOWN, SQZ_MANAGER to DOWN, and manualed ISC_LOCK to IDLE
Before the model restarts that were required, I put the corresponding suspensions into SAFE and the HAMs and BSCs into ISI_DAMPED_HEPI_OFFLINE
Once the CDS issue was sorted out, I untripped all the SUS WDs that had tripped and took the chambers back to ISOLATED/FULLY ISOLATED
13:43 Started relocking, went to GREEN_ARMS_MANUAL before starting an initial alignment
- Having issues with the ALS XARM beatnote
- Tried toggling No Force/Force - didn't work
- Not due to low beatnote
- Reset all optics except sqz to a time when we were in READY two locks ago (1396640117) - worked
14:31 INITAL_ALIGNMENT started