Reports until 09:00, Tuesday 09 April 2024
H1 SQZ
Link
daniel.sigg@LIGO.ORG -
posted 09:00, Tuesday 09 April 2024
(77049)
CLF VCO Sensitivity
Measured the sensitivity of the new CLF VCO:
Offset at Tune Input
Tune readback
Frequency (MHz)
Difference (kHz)
-1V
3.332V
202.690924
436.1
0V
4.067V
203.127060
0
+1V
4.800V
203.542136
415.1
which results in ~425kHz/V.