Displaying report 1-1 of 1.
Reports until 20:05, Monday 08 April 2024
H1 General
ryan.crouch@LIGO.ORG - posted 20:05, Monday 08 April 2024 - last comment - 10:33, Tuesday 09 April 2024(77034)
OPS Monday eve shift update

We've been locked for 6:50, range is lower tonight

A 6.0 from Indonesia rolled through around 00:38 UTC

Comments related to this report
ryan.crouch@LIGO.ORG - 21:23, Monday 08 April 2024 (77037)SQZ

SQZer lost lock and dropped us from observing at 04:22, will return once it relocks. Back into observing at 04:26UTC

camilla.compton@LIGO.ORG - 10:33, Tuesday 09 April 2024 (77054)SQZ

This lockloss was due to the OPO PZT running out of range @40V, plot. There's already a checker in SQZ_MANGER to relock when IFOs unlocked and OPO PZT1 is below 50V but it was at 70V when the IFO locked.

Images attached to this comment
Displaying report 1-1 of 1.