Reports until 16:14, Monday 15 April 2024
H1 CDS
david.barker@LIGO.ORG - posted 16:14, Monday 15 April 2024 (77187)
New SUS OPLEV Overview MEDM with links to Control Screens

Fernando, Dave:

We have released a new SUS Oplev Overview MEDM in which the BSC oplevs have CONTROL buttons to open Fernando's control screens.

I took the opportunity to macrotize the overview, with the oveview opening two MICRO MEDMs: H1SUS_OPLEV_MICRO.adl is opened directly for the HLTSs, and H1SUS_OPLEV_WCTRL_MICRO.adl is opened for the BSCs. The latter in turn opens H1SUS_OPLEV_MICRO.adl and adds the CONTROL button. The control adl file resides in the SYS/H1/MEDM directory and is called H1SYS_OPLEV_CONTROL.adl.

The new Oplev Overview was created in the SUS/H1/MEDM directory, the original remains in the SUS/COMMON/MEDM directory and has not been modified.

To see the new Oplev MEDM you will need to start a new SITEMAP.

Attachment shows the Overview MEDM, with ETMY's Control MEDM superimposed.

Images attached to this report