During yesterday's (4/16) maintenance period, a Norco tech came to the site to inspect the 8 LN2 dewars that feed the cryopumps. Inspection report will be posted to Q2000008 once received.
The vacuum jacket pressures were also measured during inspection:
| Dewar | Pressure (micron/mtorr) |
| CP1 | 68 (gauge fluctuated may need replacing, lowest value recorded) |
| CP2 | 26 |
| CP3 | 68 |
| CP4 (not in service) | 110 |
| CP5 | 9 |
| CP6 | 33 |
| CP7 | 48 |
| CP8 | 77 (gauge fluctuated may need replacing, lowest value recorded) |