After coordinating with LLO & notifying Virgo, went with running the Saturday Calibration Measurements (broadband [~5min] + simullines [~25min]) at ~ 2200utc.
LOCKLOSS (at 2225utc) During Measurement!! Sadly, it looks like we had a lockloss toward the end of the Calibration. Vlad said we should let Louis know so he could look at the script to see if the excitations are too strong toward the end of the Simulines measurement.
- Following instructions (link)
- NOTE: In the procedure there is mention of "Take ISC_LOCK to AUTO (from MANAGED by H1_MANAGER....will need to undo this when done)." But H1_MANAGER is in IDLE after we revert on-call operator at the beginning of the shift---so ISC_LOCK is already in AUTO.
- 2200 ISC LOCK to NLN_CAL_MEAS
- 2201-2206 Broadband measurement run
- Measurement at: /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/
- 2206-xxxx Simullines measurement
- gps START time: GPS: 1397685998.905604
- gps END time: GPS: 1397687180.225995 (this was given after a few ERROR lines ran after H1 lost lock.
- Files writtten to subfolders here: /ligo/groups/cal/H1/measurements