Reports until 11:52, Tuesday 30 April 2024
H1 SQZ
andrei.danilin@LIGO.ORG - posted 11:52, Tuesday 30 April 2024 (77507)
Fiber polarization aligned

Andrei, Camilla

We aligned the fiber polarization to minimize the H1:SQZ-SHG_GR_DC_POWERMON channel, following the instructions of Report 71761 on rotating the waveplates.

The reduction in monitor power is depicted in the attached screenshot (POWERMON_lesser_timespan.png), showing a decrease from 0.24 [a.u.] to 0.03 [a.u.].

The fiber currently in use (Fiber.png) is SM980-5.8-125 - Single Mode Optical Fiber, 980 - 1550 nm, Ø125 µm Cladding. As it seems from the figure POWERMON_large_timespan.png, power is unstable in the long term. We propose to change the fiber to PS-PM980 - 970 - 1550 nm PM Photosensitive Fiber, 0.12 NA, 6.6 µm MFD and to tightly fixate the optical cable on the optical table to furhter reduce polarization shift.

Images attached to this report