Translating SQZ beam for better AS_C alignment adn OMC thoughput

Naoki, Sheila, Camilla

After the output alignment shift reported in 77427, we tried to revert SQZ to the last successful SQZ-OMC scan in 77515 but when we increased throughput, AS_C alignment was bad. Today we translated the beam to get good throughput on AS_A/B and AS_C and a centered beam on AS_C with OMC DC centering loop running. This was a move of ZM4 -1000urad, ZM5 -650urad in yaw as reported by DAMP_Y_INMON channels

Attached shows previous April 16th OMC-SQZ scan compared to during today's translation and how we left the ZMs:

We're not sure how this change will effect in-lock SQZ as Vicky had to make big changed in 77400  and we may have seen that good OMC scan alignment isn't good SQZ alignment?

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 16:42, Tuesday 30 April 2024 (77530)

Naoki, Sheila, Camilla

This alignment didn't give us good in-vac SQZ.

We tried taking SQZ sliders to before the output arm alinement shift when we had these PSAMS (2024/04/17 23:38 UTC) with the addition of the -1000urad, -650urad offsets on ZM4 and ZM5. This made SQZ worse.

We moved ZM6 to maximize RF3 and RF6 adn then went to anti-sqz and ran SCAN_ALIGMENT_FDS twice. SDFs attached.

Naoki adjusted the SQZ angle to squeezing and reran SCAN_SQZANG, this got us back to 4+dB of SQZ and over 150MPc of range, still room for improvement but much better than last week.

Images attached to this comment