Minhyo
Following the changing of the gain value of the low pass filter in the range BLRMs (OAF_RANGE_RLP#, refer to alog 77346), I modified the ndscope template of 'BLRMS together'.
The original template showed the output of each RLP in one panel and BNS range (H1:CDS-SENSMON_CAL_SNSW_EFFECTIVE_RANGE_MPC and H1:CDS-SENSMON_CLEAN_SNSW_EFFECTIVE_RANGE_MPC) in the other panel. For visibility, I separated the RLP output panel into two parts (RLP_1~5 / RLP_6,7) in the left part of the whole screen and added the BNS range corresponding to each frequency band in the right part of the screen.
The template is saved in userapps/isc/h1/ndscope/DARM_BLRMS_together.yaml.
(The old one is copied as named DARM_BLRMS_together_old.yaml in the same directory)
After changing the gain value of each RLP, the outputs and the ranges for each bandwidth show better correspondence to the actual BNS range, which will be useful to see the contribution of each bandwidth range.
Screenshots are: 1) Old plot of DARM_BLRMS_together, 2) same plot with new template.