Displaying report 1-1 of 1.
Reports until 08:47, Tuesday 14 May 2024
H1 AOS
jason.oberling@LIGO.ORG - posted 08:47, Tuesday 14 May 2024 (77831)
SR3 Optical Lever Re-Centered (WP 11861)

The recent alignment shift in the IFO's output arm caused the SR3 optical lever beam to fall off of its QPD, so I re-centered it this morning.  I found the OpLev with the following readings:

The SUM counts are very low (usual is just over 20k), indicating the OpLev beam was barely on its QPD.  In fact, as I adjusted the horizontal translation stage the yaw reading went further positive to ~31 µrad before moving towards 0.  The OpLev is now re-centered with a normal SUM count reading, see attached picture.  This closes LHO WP 11861.

Images attached to this report
Displaying report 1-1 of 1.