Reports until 08:34, Friday 17 May 2024
H1 PSL
ryan.crouch@LIGO.ORG - posted 08:34, Friday 17 May 2024 (77883)
PSL Status Report - Weekly

Closes FAMIS 26244

PMC refl has been trending up over the past week, I've also been getting "ISS diff power is low" messages from diag_main.
Laser Status:
    NPRO output power is 1.818W (nominal ~2W)
    AMP1 output power is 66.81W (nominal ~70W)
    AMP2 output power is 137.3W (nominal 135-140W)
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN
    PDWD watchdog is GREEN

PMC:
    It has been locked 44 days, 21 hr 29 minutes
    Reflected power = 19.05W
    Transmitted power = 107.7W
    PowerSum = 126.7W

FSS:
    It has been locked for 0 days 12 hr and 49 min
    TPD[V] = 0.9025V

ISS:
    The diffracted power is around 2.3%
    Last saturation event was 0 days 12 hours and 49 minutes ago


Possible Issues:
    PMC reflected power is high

Images attached to this report