H1 TCS
camilla.compton@LIGO.ORG - posted 16:31, Thursday 29 February 2024 - last comment - 00:34, Monday 20 May 2024(76056)
Measured Distance between ALS table and Viewport

At LLO, Aidan and Karla found the ISCTEX/Y tables (ALS/HWS) tables were further from the BSC viewport than designed which would change Aidan's T1000717  mode-matching calculations. Ours are also different, EY more than EX. 

Distances are from the ISC table horizontally to the viewport outside edge, as in D0902168 and D0902163.
Measured to the back edge of VP closest to chamber and removed 2" for VP flange thickness, in photos the tape measure distance is 3". 
 
  Design Measured (+/-0.5") Difference 
ISCTEX to BSC10 right VP 16.5" D0902168  19.5" photo 3" = 75mm
ISCTEY to BSC9 left VP 16.5" D0902163 24.5"  photo 8" = 200mm
Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 00:34, Monday 20 May 2024 (77936)

Madison Simmonds, Camilla Compton

G1100873-v4 shows this center of VP to table distance is 18.6" which is closer to what was measured but not what was used in the HWS calculations.