naoki.aritomi@LIGO.ORG - posted 14:37, Thursday 23 May 2024 - last comment - 18:09, Tuesday 28 May 2024(78000)
NLG sweep on DARM

Naoki, Karmeng, Andrei, Sheila

We did NLG sweep on DARM at NLG 16.9 (nominal) and 42. The IFO was locked for more than 20 hours and well thermalized. We also tried NLG of 65.9, but the squeezing level was not stable so we gave up this NLG. We will take the third NLG data tomorrow if IFO is thermalized. 

Previous NLG sweep on DARM: 73747

How to measure NLG: 76542

How to change NLG: 73801

Note:

  UTC demod phase DTT ref NLG SQZ at 2kHz (dB)
FDS 16:29:00-16:34:00 160.91 3 16.9 -4.2
No SQZ 16:36:20-16:41:20   0    
ASQZ 16:56:20-17:01:20 -100.37 6 16.9 15.4
mean SQZ 17:11:00-17:16:00   7 16.9 12.6
FDS 18:02:10-18:07:10 162.13 4 42 -4.4
ASQZ 18:19:30-18:24:00 -87.6 8 42 20
mean SQZ 18:29:00-18:34:00   9 42 17

 

OPO trans (uW) OPO temp Seed amplified Seed unamplified NLG
80 31.470 0.215 0.0127 16.9
100 31.454 0.534 0.0127 42
110 31.448 0.0725 0.0011 65.9

 

Images attached to this report
Comments related to this report
naoki.aritomi@LIGO.ORG - 12:32, Friday 24 May 2024 (78025)

After PR2 spot move yesterday in 78012, we did NLG sweep again. This time we took three NLG at 7.9, 16.3 (nominal), 55.9.

  UTC demod phase DTT ref NLG SQZ at 2kHz (dB)
No SQZ 16:41:00-16:46:00   0    
FDS 16:54:08-16:59:08 161.59 10 7.9 -4.4
ASQZ 17:04:00-17:09:00 -95.63 11 7.9 11.1
mean SQZ 17:10:52-17:15:52   12 7.9 9.1
FDS 17:31:40-17:36:40 192.47 13 55.9 -4.1
ASQZ 17:43:54-17:48:54 -86.42 14 55.9 21.2
mean SQZ 17:50:53-17:55:53     55.9 18.5
FDS 18:12:31-18:17:00 190.31 15 16.3 -4.7
ASQZ 18:21:54-18:26:54 -97.79 16 16.3 15.3
mean SQZ 18:28:12-18:33:12   17 16.3 12.5

 

OPO trans (uW) OPO temp Seed amplified Seend unamplified NLG
60 31.484 0.0086 0.00109 7.9
80 31.468 0.0174 0.00107 16.3
105 31.446 0.0609 0.00109 55.9

 

Images attached to this comment
victoriaa.xu@LIGO.ORG - 18:09, Tuesday 28 May 2024 (78097)

Vicky, Karmeng

This NLG scan is compatible with ~30% SQZ losses, ~20 mrad phase noise.

Attachment 1, 2 - Calculated loss ~30% from mean sqz and generated sqz, then fit ASQZ/SQZ to estimate phase noise ~ 20 mrad and technical noise. If fitting SQZ+ASQZ together to estimate, then fit loss ~32-33%. This uses standard linear opo equations to estimate generated squeezing level based on NLG.

Attachment 3 - Calculated loss ~ 27% from mean sqz and generated sqz, then fit ASQZ/SQZ to estimate phase noise ~ 20 mrad and technical noise. This uses bowtie opo equations to estimate generated squeezing level based on NLG (few % lower generated sqz than the above estimate). See Eq. 13 of Dhruva's ADF paper, e.g. P2200041.

NLG calibration - Estimates OPO green pump trans threshold @ 142 uW. This seems close to previous threshold estimates ~149 uW made just after moving to this crystal spot (LHO:73562, Oct2023 crystal move).

A comparison of O4a (hd,ifo) and O4b NLG (ifo) scans, maybe most interesting is comparing homodyne vs. interferometer squeezing in O4a (~32% loss, LHO:78000). O4a/O4b IFO losses look similar, but I think it's largely an issue with this O4b measurement?

 

A note about this NLG scan - I think total optical losses should be < 30% based on seeing >5dB SQZ previously. For example, the -5.4 dB SQZ observed in LHO:76553 is too much squeezing, and incompatible with losses >30%. So I think this measurement has higher losses than "normal" in O4b, maybe related to the alignment / mode-matching / (something that drifts) not being optimal here. Would be interesting to get back to the -5dB spot (of course), and see how losses look then.

 