PSL Weekly Status Update Famis 26245
Laser Status:
NPRO output power is 1.819W (nominal ~2W)
AMP1 output power is 66.51W (nominal ~70W)
AMP2 output power is 137.7W (nominal 135-140W)
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 54 days, 21 hr 43 minutes
Reflected power = 19.44W
Transmitted power = 107.0W
PowerSum = 126.4W
FSS:
It has been locked for 0 days 17 hr and 35 min
TPD[V] = 0.89V
ISS:
The diffracted power is around 2.6%
Last saturation event was 0 days 17 hours and 34 minutes ago
Possible Issues:
PMC reflected power is high