Jim, Sebastien
We made good progress on the ISI-ETMX testing today. All the sensors have been tested and react properly. Actuator side, we noticed an unwanted ouput voltage coming out of the corner 3 coil driver. We should be able to fix this issue with Richard tomorrow. Except corner 3, all the actuators behave properly.
I've started a set of L2L measurements for corner 1 and 2 tonight. If everything goes well, I'll complete this set tomorrow night.
Tonight measurement should be done by 6am tomorrow morning.