Reports until 12:26, Tuesday 09 July 2024
H1 PEM
ryan.crouch@LIGO.ORG - posted 12:26, Tuesday 09 July 2024 (78351)
Dust monitor FAMIS testing

All the Dust Monitors in use passed the zero count and flow rate test for this cycle. Closes FAMIS27659, this was last done by me last December alog74610.

I ran the zero count and flow rate test for all the dust monitors in use around the site. Passing is having zero counts of 0.3 and 0.5 micron particles and a flow rate of 2.8 L/min.

We also have 4 working spares, 3 pumpless & 1 pumped, 3 (the 3 PL) of which have been calibrated by METONE earlier this year.

CS:

ARMs:

EX: Passed zero count on the first test and flow rate test.

EY: 2 0.3 particle counts on the first test, zero on the second which is a pass, and the flow rate was good.