Displaying report 1-1 of 1.
Reports until 09:41, Monday 10 June 2024
H1 ISC
jennifer.wright@LIGO.ORG - posted 09:41, Monday 10 June 2024 - last comment - 16:41, Wednesday 12 June 2024(78342)
Ran A2L scripts to reduce A2L coupling

Jennie W, Sheila, TJ

 

This is done to reduce the coupling we get to length from having off-centre spot positions on the mirrors.

The script can be found in userapps/isc/common/scripts/decoup/a2l_min_generic_LHO.py .

See the steps the script runs in the bottom two plots and the reponse of the ADS demodulated signals in the top plots.

I ran ETMX P and Y first by calling the script with

python a2l_min_generic_LHO ETMX BOTH

changed P and Y

H1:SUS-ETMX_L2_DRIVEALIGN_P2L_SPOT_GAIN => 3.22

H1:SUS-ETMX_L2_DRIVEALIGN_Y2L_SPOT_GAIN => 4.99

 

ETMY

both P and Y changed

H1:SUS-ETMY_L2_DRIVEALIGN_P2L_SPOT_GAIN => 4.3

H1:SUS-ETMY_L2_DRIVEALIGN_Y2L_SPOT_GAIN => 0.92

 

ITMX

both P and Y changed

H1:SUS-ITMX_L2_DRIVEALIGN_P2L_SPOT_GAIN => -0.99

H1:SUS-ITMX_L2_DRIVEALIGN_Y2L_SPOT_GAIN => 2.88

 

ITMY

H1:SUS-ITMY_L2_DRIVEALIGN_P2L_SPOT_GAIN => -0.36

H1:SUS-ITMY_L2_DRIVEALIGN_Y2L_SPOT_GAIN => -2.5
 

16:35:52 UTC all A2L gains set and script finished.


 

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 10:38, Monday 10 June 2024 (78347)

CHARD_Y coupling was still bad so we reran Y degree of freedom for all test masses.

ETMX

Want to change gain from 4.99 to 4.98, rounded to 2 decimal places. St.Div is 0.925. Change of 0.01
H1:SUS-ETMX_L2_DRIVEALIGN_Y2L_SPOT_GAIN => 4.98

 

ETMY

Want to change gain from 0.92 to 0.93, rounded to 2 decimal places. St.Div is 0.145. Change of -0.007
H1:SUS-ETMY_L2_DRIVEALIGN_Y2L_SPOT_GAIN => 0.93

 

ITMY

Want to change gain from -2.5 to -2.5, rounded to 2 decimal places. St.Div is 1.0. Change of -0.001
Gain did not change.

3 A2Ls changed by 17:33:32 UTC.

ITMX

Want to change gain from 2.88 to 2.87, rounded to 2 decimal places. St.Div is 0.496. Change of 0.01
H1:SUS-ITMX_L2_DRIVEALIGN_Y2L_SPOT_GAIN => 2.87


3 A2Ls changed by 17:33:32 UTC.

 

jenne.driggers@LIGO.ORG - 16:41, Wednesday 12 June 2024 (78402)

I don't think these values made it into lscparams' dictionary a2l_gains['FINAL'] (starting around line 451 in lscparams.py), so they were reverted for the next lock, and they were accepted as the previous values in alog 78383.

Displaying report 1-1 of 1.