Displaying report 1-1 of 1.
Reports until 16:52, Tuesday 18 June 2024
H1 SQZ
naoki.aritomi@LIGO.ORG - posted 16:52, Tuesday 18 June 2024 (78529)
Optimal SQZ angle depends on OPO PZT voltage

As shown in the attachment, SQZ unlocked for unknown reason at 2024/6/18 23:12:09 UTC and the range got much worse after it relocked. We ran the SCAN_SQZANG and the range was recovered. The optimal SQZ angle changed from 225 to 203 after relocking and it seems due to the different OPO PZT voltage.