H1 ISC
francisco.llamas@LIGO.ORG - posted 17:27, Wednesday 03 July 2024 (78606)
Measurement of sensing function under DHARD changes

LouisD, SheilaD, FranciscoL

On Thursday, June 13 2024, we made simuline measurements after changing DHARD pitch and yaw gains. We see a change in the sensing function from a 50% increase of the DHARD pitch and yaw gains.

Routine calibration measurements done at 2024-06-13 15:46 UTC (78409). We increased (in magnitude) H1:ASC-DHARD_Y_GAIN from -40 to -60, and H1:ASC-RPC_DHARD_Y_GAIN from -30 to -45 -- a 50% increase -- and ran a second calibration measurement at 2024-06-13 16:56 UTC. The trend dhard_gain.png shows the values from these channels within the calibration measurements time window. The DHARD gains were reverted around 2024-06-13 17:40 UTC.

Plotting the transfer function of both measurements on out.png shows a significant change in the sensing function in the low frequency (<10 Hz). The uncertainty is below 2% for most of both measurements as seen in unc.png. These plots were produced using compare_measurements.py (full path: /ligo/home/francisco.llamas/CAL/sensing_function/20240621_plot_sensing_measurements/compare_measurements.py). We had to run git config --global safe.directory='*' where the '*' should be the changed by the name of the report that the script was unable to find.

Follow up analysis and plots to come.