Displaying report 1-1 of 1.
Reports until 14:11, Monday 24 June 2024
H1 AOS (ISC, OpsInfo)
jennifer.wright@LIGO.ORG - posted 14:11, Monday 24 June 2024 (78628)
Ran A2l, accepted new values in OBSERVE.snap

Jennie, TJ

 

I ran TJ's script to measure and set all the A2L gains for the quads.

16:05 UTC ran script in userapps/isc/h1/scripts/a2l
 

python a2l_min_multi.py -a
 

*************************************************
RESULTS
*************************************************
          ETMX P
Initial:  3.26
Final:    3.19
Diff:     -0.07

 

          ETMX Y
Initial:  4.89
Final:    4.81
Diff:     -0.08

 

          ETMY P
Initial:  4.35
Final:    4.39
Diff:     0.04

 

          ETMY Y
Initial:  1.06
Final:    1.13
Diff:     0.07

 

          ITMX P
Initial:  -1.0
Final:    -1.02
Diff:     -0.02

 

          ITMX Y
Initial:  2.8
Final:    2.73
Diff:     -0.07

 

          ITMY P
Initial:  -0.38
Final:    -0.41
Diff:     -0.03

 

          ITMY Y
Initial:  -2.41
Final:    -2.28
Diff:     0.13

16:14 UTCscript finished.

TJ added these to lscparams.py and so they were loaded when Corey re-locked after the commissioning period (around 20:50 UTC).

I accepted these in OBSERVE.snap, see attached photos.

Images attached to this report
Displaying report 1-1 of 1.