francisco.llamas@LIGO.ORG - posted 14:53, Tuesday 02 July 2024 (78807)
Pcal XY comparison investigation - seventh movement, inner beam up

RickS, FranciscoL

After one week of having the inner beam centered on the Rx sensor (alog 78514), on June 25, we moved the beam from center to up.

The last page of EndStationLog.pdf shows a diagram of the movements that have been done and, on dashed lines, the move that were left to do. The previous to last page on file lists the voltage values after each significant step of procedure T2400163. The steps represent writing down a voltage value after a particular change to the beam position. Some highlights from the recorded values.

The 'Initial' measurement changed by 11% from the last voltage measurement from the previous movement, done on June 18 (alog 78514). The initial and final voltage measurement during today procedure changed by 0.03%.

Images attached to this report