H1 SQZ (SQZ)
ryan.crouch@LIGO.ORG - posted 01:00, Friday 05 July 2024 - last comment - 12:37, Friday 05 July 2024(78867)
OPS Thursday eve shift summmary

TITLE: 07/05 Eve Shift: 2300-0800 UTC (1600-0100 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: Ryan S
SHIFT SUMMARY: An earthquake lockloss then a PI lockloss. Currently at MAX_POWER

Lock1:
Lock2:

Lock3:

 

To recap for SQZ, I have unmonitored 3 SQZ channels on syscssqz (H1:SQZ-FIBR_SERVO_COMGAIN, H1:SQZ-FIBR_SERVO_FASTGAIN, H1:SQZ-FIBR_LOCK_TEMPERATURECONTROLS_ON) that keep dropping us out of observing for now until their root issue can be fixed (Fiber trans PD error, too much power on FIBR_TRANS?). I noticed that each time the GAINS change it also drops our cleaned range

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 12:37, Friday 05 July 2024 (78887)

It seems that as you found, the issue was the max power threshold. Once Ryan raised the threshold in 78881, we didn't see this happen again, plot attached. I've re-monitored these 3 SQZ channels: sdfs attached (H1:SQZ-FIBR_SERVO_COMGAIN, H1:SQZ-FIBR_SERVO_FASTGAIN, H1:SQZ-FIBR_LOCK_TEMPERATURECONTROLS_ON) and TEMPERATURECONTROLS_ON accepted.

It's expected that the CLEAN range would drop as that range only reports when the GRD-IFO_READY flag is true (which isn't the case when there's sdf diffs).

Images attached to this comment