jeffrey.bartlett@LIGO.ORG - posted 11:55, Friday 27 September 2013 (7888)
Check out ETM-X and TMS-X
   I pulled up the covers on ETM-X and TMS-X to examine the suspensions after the cartridge installation in BSC9. Both suspensions appear to have made the trip in fine shape. All the fibers are intact, the masses have not shifted on the EQ stops, and all suspension wires are intact. 

   Both suspensions are still on their stops and covered in the C3 socks. 

   Note: The lower left flag on the Penultimate mass is very close to the AOSEM on the Pen-Re mass. They are not touching. Extra care will be needed when releasing these masses from the EQ stops to ensure they do not touch.        