david.barker@LIGO.ORG - posted 01:12, Saturday 06 July 2024 - last comment - 09:26, Saturday 06 July 2024(78892)
Timing Error, restart of SUS frontends needed

Tony, Jim, Erik, Dave:

We had a timing error which caused DACKILLs on h1susb123, h1sush34, h1sush56 and DAC_FIFO errors on h1sush2a.

There was no obvious cause of the timing error which caused the Dolphin glitch, we noted that h1calcs was the only model with a DAQ CRC error (see attached).

After diag-reset and crc-reset only the SUS dackill and fifo error persisted. 

We restarted all the models on h1susb123, h1sush2a, h1sush34, h1sush56 after bypassing the SWWD SEI systems for BSC1,2,3 and HAM2,3,4,5,6.

SUS models came back OK, we removed the SEI SWWD bypasses and handed the system over to Tony.

Images attached to this report
Comments related to this report
erik.vonreis@LIGO.ORG - 01:46, Saturday 06 July 2024 (78894)

H1SUSH2A DACs went into error 300 ms before the CRC SUM increased on h1calcs. 

 

The DAQ (I believe) reports CRC errors 120 ms after a dropped packet, leaving 180 ms. unaccounted for.

Images attached to this comment
david.barker@LIGO.ORG - 09:17, Saturday 06 July 2024 (78901)

FRS31532 created for this issue. I has been closed as resolved-by-restarting.

david.barker@LIGO.ORG - 09:26, Saturday 06 July 2024 (78902)

Model restart logs from this morning:

Sat06Jul2024
LOC TIME HOSTNAME     MODEL/REBOOT
01:12:22 h1susb123    h1iopsusb123
01:12:33 h1sush2a     h1iopsush2a 
01:12:39 h1sush34     h1iopsush34 
01:12:43 h1susb123    h1susitmy   
01:12:47 h1sush2a     h1susmc1    
01:12:56 h1sush56     h1iopsush56 
01:12:57 h1susb123    h1susbs     
01:12:59 h1sush34     h1susmc2    
01:13:01 h1sush2a     h1susmc3    
01:13:10 h1sush56     h1sussrm    
01:13:11 h1susb123    h1susitmx   
01:13:13 h1sush34     h1suspr2    
01:13:15 h1sush2a     h1susprm    
01:13:24 h1sush56     h1sussr3    
01:13:25 h1susb123    h1susitmpi  
01:13:27 h1sush34     h1sussr2    
01:13:29 h1sush2a     h1suspr3    
01:13:38 h1sush56     h1susifoout 
01:13:52 h1sush56     h1sussqzout 
 

Displaying report 1-1 of 1.