H1 PEM
robert.schofield@LIGO.ORG - posted 16:22, Sunday 07 July 2024 (78925)
Slow bias changes during observation completed; New ESD bias settings: ITMX 0V, ITMY -40V (-4.0 in offset window)

Commissioning time has been reduced recently by lock losses and other issues, so I have been trying a new technique for determining the ITM biases that minimize coupling of electronics ground noise. I have been making slow adjustments of the bias during observation, recording the times of those slow changes, and then monitoring the coherence between DARM and the current clamp on the electronics grounding cable in the CER for the new settings.

The times of the slow changes during observation mode, and the bias setting I have tried, are listed in 78791, 78767, 78734 and this log. The figure compares the original settings for O4b, the setting that we guessed would be best, 0V,0V, and the setting that was consistently better, ITMX: 0, ITMY: -40 V (a setting of -4 in the offset window).

Start of change (GPS)

End of change (GPS)

ITMX bias at start (V)

ITMX bias at end (V)

ITMY bias at start (V)

ITMY bias at end (V)

1404317823

1404317979

0

-20

0

-22

1404319278

1404319453

-20

20

-22

-20

1404341414

1404341526

0

0

0

-20

1404342991

1404343117

0

0

-20

-40

1404344533

1404344655

0

40

-40

-40

1404413221

1404413147

0

0

0

-40

Non-image files attached to this report