Laser Status:
NPRO output power is 1.821W (nominal ~2W)
AMP1 output power is 65.27W (nominal ~70W)
AMP2 output power is 138.5W (nominal 135-140W)
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 2 days, 23 hr 37 minutes
Reflected power = 17.94W
Transmitted power = 107.8W
PowerSum = 125.8W
FSS:
It has been locked for 0 days 0 hr and 49 min
TPD[V] = 0.7949V
ISS:
The diffracted power is around 3.2%
Last saturation event was 0 days 0 hours and 49 minutes ago
Possible Issues:
ISS diffracted power is high, seems like its been high since early morning this past maintenance.