david.barker@LIGO.ORG - posted 09:15, Sunday 14 July 2024 - last comment - 09:33, Monday 15 July 2024(79108)
Pressure glitch on PT100B HAM1 at 07:25:15 Sun 14 July 2024

There was a pressure glitch on PT100B HAM1 cold-cathode at 07:25:15 PDT this morning. There was no corresponding glitch on any other corner station gauge as far as I can tell.

Pressure increased from 3.3e-08 to 4.0e-08 Torr, recovered in 4 minutes.

Images attached to this report
Comments related to this report
jenne.driggers@LIGO.ORG - 11:31, Sunday 14 July 2024 (79109)

This is clearly very concerning. 

Attached is a trend of the pressure in the top right, along with several other channels (IMC power in, ISC_LOCK guardian state, RM alignments, DC power levels on all 4 PDs in that chamber), and I don't see anything in any of these channels that looks suspicious.

EDIT to add second screenshot, of in-vac tabletop L4C seismic sensors.  Again, nothing suspicious-looking to me that I see yet.

Images attached to this comment
michael.zucker@LIGO.ORG - 07:41, Monday 15 July 2024 (79112)

Given no lockloss correlation, may not be significant; ~3 orders of magnitude less PV than the June event(s).   Maybe an ion pump burped up some argon.

jordan.vanosky@LIGO.ORG - 09:33, Monday 15 July 2024 (79121)

IP13 ion current/pressure spiked up ~1 second before the PT100B gauge did, see attached. Likely argon instability of the pump

Images attached to this comment