SR3 transfer functions

The seismic state of HAM5 during this measurement was, ISI_DAMPED_HEPI_OFFLINE and the HEPI was locked yesterday. I put SR3 to DAMPED, made sure the damping outputs were off and the TEST outputs were on. Overall it looks fairly close to the model.

On Roll I saw an extra peak around ~1.04 Hz, and Verticle around ~1.5 Hz (cross coupling from yaw?).

The templates are saved at /ligo/svncommon/SusSVN/sus/trunk/HLTS/H1/SR3/SAGM1/Data

Non-image files attached to this report
Comments related to this report
ibrahim.abouelfettouh@LIGO.ORG - 16:36, Wednesday 17 July 2024 (79201)
ryan.crouch@LIGO.ORG - 11:19, Thursday 18 July 2024 (79217)