francisco.llamas@LIGO.ORG - posted 22:17, Wednesday 24 July 2024 - last comment - 10:18, Thursday 01 August 2024(79300)
HAM 5 laser alignment day 3

Camilla, Jason, Jenne, Keita, Sheila, Francisco, Naoki

Summary: We are done for AUX laser alignment and power measurement of AUX laser. SQZ beam got to OM1, but we did not do power measurement due to noisy seed dither lock. The following are a mixture of notes between Camilla, Francisco, and Naoki regarding the changes done today:

Aiming for 2 1/4" above EQ stop as 4" between vertical stops and the center of SR2 optic is 5.5mm above center line of EQ stops.

At start of the day, someting had drifted (mainly pitch), re-centered AUX laser on irises.

*Iterating moving auxiliary laser mirror and beamspiltter to center in om1 and retroreflection irises.*

Checking position at SR2:

Swapped the AUX laser steeing mirror to a better optic mount and not woberly perdistal, we are blaming this for our drift.

Jenne also adjusted SRM sliders to yesterdays values.

*Iterating moving auxiliary laser mirror and beamspiltter to center in om1 and retroreflection irises.*

Checking position at SR2:

Moved SRM osems from (175, 140) to (50,190). In pitch moved -125urad from 175 to 50 on osems. In yaw moved +50urad. this was 140 to 190 on osems
Aiming to center the beam on SR2


*Iterating moving auxiliary moving mirror to center in om1 and retroreflection irises.*

Checking position at SR2:

EQ stops are 2 1/2 in away from each other; center is 1 1/4 in from horizontal EQ stop

Will do 100 microrad in -YAW, then check retro in OM1. -100urad in yaw is 190 to 90 on osems

*Iterating moving auxiliary moving mirror to center in om1 and retroreflection irises.*

Checking positions at SR2:

In yaw going +300 urad. This is 90 to 390 on osems.

*Iterating moving auxiliary moving mirror to center in om1 and retroreflection irises.*

Checking positions at SR2:

*LUNCH BREAK*
 


Sheila back to SR2;

earlier today SRM P moved from 179 to 50. 1 1/4 inch move on SR2

Moved 1/4 inch backward. moved SRM P from 50 to 76

Sheila is happy with SR2 centering.

Jason is setting iris on OM1. Beam on OM1 ugly? due to crater on OFI?

ASC_AS_C P -0.6 Y -0.6

open GV for HAM7. send SQZ beam. seed dither lock is very noisy.

SQZ beam is close to AUX laser?

AUX power measurement

Comments related to this report
camilla.compton@LIGO.ORG - 13:49, Thursday 25 July 2024 (79313)

First three attached photos are of the aux laser steering mirror we swapped in, original mirror in 79253.

Fourth photo is the iris we places between the aux laser beamsplitter and SRM. Once we were finished with the alignment Jenne and Sheila recentered this iris on the beam. 

The power measurements recorded are with the AUX laser as we had trouble dither locking the SQZ seed beam. More details:

  • 1.72mW before SRM (SR2 side)
  • 0.605mW just after SRM (Good, expect 0.602mW =  35% transmitted = 0.35 x 1.72mW)
  • 0.45mW after KTP and TGG before TFP (expect 100% throughput so are loosing 25% through OFI)
  • 0.380mW before OM1 (some power lost through wring polarization to OFI rejected path? Didn't confirm).
Fifth image is the SQZ beam after passing through the OFI and SRM. Can see a big whole in the middle of the beam!! We expect this is from passing through the KTP crater! Still not easy to tell which crater the beam is passing though.
Note that the aux laser beam is too small to tell by eye if there is a hole in it, but once we got the alignment better, the beam quality seemed to decrease and the beam become dimmer.

Sixth image is the SQZ beam and aux laser beam (smaller) after SRM. They are misaligned in yaw. 

Seventh image is te  SQZ beam and aux laser beam (below) before OM1. They are misaligned in pitch.

Images attached to this comment
corey.gray@LIGO.ORG - 10:18, Thursday 01 August 2024 (79402)EPO

Tagging for EPO.