Reports until 17:01, Tuesday 01 October 2013
H1 SEI
hugh.radkins@LIGO.ORG - posted 17:01, Tuesday 01 October 2013 (7944)
WBSC9 ETMX In-Air at-Chamber Cabling secured--CPSs Grounded
The CPS on board target ground lugs are tied to chamber ground.  All sensor cables are strain relieved at the feedthru protection shrouds.  The EM actuator cables still need to be strain relieved.